Specifications
- Output Signal
- 4-20 mA
- Rangedown
- Up To 10:1
- Measurement Range
- Up to 2000 psi (13.79 MPa) Differential, Up to 4000 psig (27.58 MPa) Gage
- Housing Material
- SST
- Process Wetted Material
- 316L SST
- Seismic Rating
- 8.5g ZPA Seismic
- Radiation Rating
- 165 Mrad (1.65 kGy) TID Gamma Radiation
- Steam/Pressure Temperature
- 437°F (225°C) Steam/Temperature
- Qualification Standard
- IEEE Std 323™-1974/1983/2003
IEEE Std 344™ – 1975/1987/2004
- Product Literature
- The Product Data Sheets and Manuals posted on this web-site are intended for general product information only, with errata included. They may not represent the latest revision or contain information relating to recent product notifications, if any.
- Contact Information
- Please contact Rosemount Nuclear Instruments, Inc. at 1-952-949-5200 or RNII.info@emerson.com for the most current information available
Features
- Qualified per IEEE Std 323™-1974/1983/2003 for an industry-standard solution
- Delivers 0.2% reference accuracy for reliable measurements
- Withstands radiation exposure to 167 Mrads TID gamma and seismic levels to 8.5g ZPA
- Durable design engineered for steam temperature up to 437 °F (225 °C) performance